Socket design and cement pressurisation in the Charnley low-friction arthroplasty |
| |
Authors: | P Shelley B M Wroblewski |
| |
Affiliation: | Centre for Hip Surgery, Wrightington Hospital, Wigan, Lancashire, England. |
| |
Abstract: | An experimental evaluation was made of the efficacy of an unflanged cup, an ogee-flanged cup and the Exeter pressuriser in the pressurisation of acetabular cement. Only a very modest injection pressure could be generated through an unflanged socket and even this was readily lost as the socket "bottomed out". The ogee-flanged socket gave a consistently high injection pressure which could be maintained throughout the process of polymerisation. The Exeter pressuriser gave marginally better results. Of the two cements tested, Palacos with gentamicin showed better intrusion than CMW Type 1 cement, but it also allowed greater extrusion once pressure was released. The importance of maintaining a continuous pressure on the cement throughout polymerisation is emphasised. |
| |