首页 | 本学科首页   官方微博 | 高级检索  
     

可调恒压式毛细管粘度计误差分析
引用本文:秦任甲. 可调恒压式毛细管粘度计误差分析[J]. 中国医学物理学杂志, 1996, 13(3): 165-166
作者姓名:秦任甲
作者单位:桂林医学院血液流变学研究室
摘    要:本文介绍了可调恒压式毛细管粘度计的工作原理,详尽分析了弯液面附加压强和驱动压差不稳定引起的误差及减少误差的方法。对有关检测人员和医师正确使用这种粘度计和科学地分析,使用测量结果有指导意义。

关 键 词:毛细血管粘度计,血液粘度,误差分析

ANALYSIS OF ERROR CAUSED BY ADJUSTABLE CAPILLARY VISCOSIMETER WITH CONSTANT PRESSURE
Qin Renjia. ANALYSIS OF ERROR CAUSED BY ADJUSTABLE CAPILLARY VISCOSIMETER WITH CONSTANT PRESSURE[J]. Chinese Journal of Medical Physics, 1996, 13(3): 165-166
Authors:Qin Renjia
Abstract:The Principle of adjustable capillary viscosimeter with constant pressure was introduced.De-tailed analysis was done as to the error caused by additive pressure of bending liquid surface and instability ofdriving pressure difference as well as the method to decrease the error.It is of guiding significance to laboratoryworders and doctcrs who can thereby perform the meter correctly and anglyse and use the results measured in ascientific way.
Keywords:Capillary viscosimter Blood viscosity Analysis of error
本文献已被 CNKI 维普 等数据库收录!