首页 | 本学科首页   官方微博 | 高级检索  
检索        


Exact measurement of vascular lumina in digital subtraction arteriography (DSA) during stent implantations and percutaneous transluminal angioplasty (PTA)
Authors:T Pfluger  G Küffer  D Hahn
Institution:(1) Department of Radiology, University of Munich, Pettenkoferstr. 8 a, W-8000 Munich 2, Germany
Abstract:There are inherent difficulties in using high-resolution DSA in interventional radiology for measuring vascular luminal diameter since enlargement ratios are not constant. We present a system that permits accurate determination of vascular lumen diameters by means of a special angiographic catheter with pre-defined marking points. The accuracy of this measuring method was explored in tests using phantoms. In vivo practicability of the method was proved in PTA and stent implantations in 39 patients. This method with a maximum error of measurement of 5% appears to provide adequate safety in the choice of suitable dilating balloons and stents in interventional vascular procedures. In follow-up, it permits the accurate measurement of neo-intimal thickness.
Keywords:Digital subtraction arteriography  Quantitative measurements  Calibration catheter
本文献已被 SpringerLink 等数据库收录!