Subcutaneous implantable cardioverter-defibrillator electrode fracture: Follow-up,troubleshooting, and evaluation
Authors:Danesh K. Kella MBBS  FHRS  Bruce S. Stambler MD  FHRS
Affiliation:Division of Clinical Cardiac Electrophysiology, Piedmont Heart Institute, Atlanta, Georgia, USA
Abstract:The subcutaneous-implantable cardioverter-defibrillator (S-ICD) and its electrode were developed to avoid long-term complications of transvenous leads in the vasculature. We report a case of unexpected, inappropriate S-ICD shocks due to oversensing of high-amplitude, nonphysiologic, electrical noise artifacts that were not preceded by high-impedance alerts or sensing electrogram noise detections. Following explant, high-magnification X-ray imaging of the S-ICD electrode demonstrated partial fracture of the distal sensing conductor located near a short radius bend in the electrode at the electrode-header interface. Clinicians should be aware of a potential for fatigue failure fracture of the S-ICD electrode. Recommendations for systematic S-ICD follow-up and troubleshooting are discussed.
Keywords:electrode fracture  inappropriate ICD therapy  oversensing  subcutaneous ICD