首页 | 本学科首页   官方微博 | 高级检索  
     


Hydrodynamic characterisation of ventricular assist devices
Authors:Vandenberghe S  Segers P  Meyns B  Verdonck P
Affiliation:IBITECH, Ghent University, Gent, Belgium. stijn.vandenberghe@navier.rug.ac.be
Abstract:A new mock circulatory system (MCS) was designed to evaluate and characterise the hydraulic performance of ventricular assist devices (VADs). The MCS consists of a preload section and a multipurpose afterload section, with an adjustable compliance chamber (C) and peripheral resistor (Rp) as principal components. The MCS was connected to a pulse duplicator system for validation, simulating a wide range of afterload conditions. Both pressure and flow were measured, and the values of the different components calculated. The data perfectly fits a 4-element electrical analogon (EA). The MCS was further used to assess the hydrodynamic characteristics of the Medos VAD as an example of a displacement pump. Data was measured for various MCS settings and at different pump rates, yielding device specific pump function graphs for water and pig blood. Our data demonstrate (i) flow sensitivity to preload and afterload and (ii) the effect of test fluid on hemodynamic performance.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号