Comparison of horizontal corneal diameter measurements using Galilei,EyeSys and Orbscan II systems |
| |
Authors: | Ramin Salouti MD Mohammad H Nowroozzadeh MD Mohammad Zamani MD Maryam Ghoreyshi MD Rosa Salouti BD |
| |
Affiliation: | Poostchi Ophthalmology Research Center, Shiraz University of Medical Sciences, Shiraz, Iran E‐mail: saloutir@hotmail.com |
| |
Abstract: | Background: White‐to‐white corneal measurement is used frequently in cataract and refractive surgery as well as in diagnosing various corneal diseases and monitoring congenital glaucoma. In this article, we determine agreement in horizontal corneal diameter measurements in normal eyes among the Galilei V4.01, the EyeSys and the Orbscan IIz. Methods: In this prospective study, 37 consecutive volunteers were enrolled. All 74 eyes were measured with all three devices. Galilei, EyeSys and Orbscan II measurements were compared with one another. In addition, the data from the right and left eyes were compared for each device. Results: The mean horizontal corneal diameter measurements with the Galilei, the EyeSys and the Orbscan II were 12.01 ± 0.61 mm, 12.09 ± 0.87 mm and 11.67 ± 0.29 mm, respectively. The difference between the Galilei and EyeSys systems was not statistically significant (p = 0.76), however, the Orbscan II showed smaller measurements (p < 0.001). The best 95% limits of agreement between devices were for the Galilei and the Orbscan II (‐0.72, 1.48; r = 0.4, p < 0.01). The best 95% limits of agreement between two eyes for each device were found with the Orbscan II (‐0.15, 0.17; r = 0.996, p < 0.01). There was no correlation between measurements of right and left eyes for the Galilei or the EyeSys (r = 0.274, p = 0.176 and r = 0.31, p = 0.065, respectively). Conclusions: These results suggest that measurements made with the Orbscan II are smaller than those obtained with the EyeSys Corneal Analysis system and the Galilei. Among the three devices, the Galilei and the Orbscan II showed the best agreement, however, it is inadvisable to use the three devices interchangeably in every clinical situation. |
| |
Keywords: | EyeSys Galilei horizontal corneal diameter Orbscan II |
|
|