Management of Atrial Tachyarrhythmias: Benefits of Pacemaker Diagnostics
Authors:LAURENT FAUCHIER  FLORENT BRIAND†  FRANÇOIS XAVIER SOTO‡  FRANCK QUENNELLE§  JACQUES LÉVY&#;  JEAN PHILIPPE DARMON#  DAVID LELLOUCHE  THOMAS LAVERGNE††  PHILIPPE PORET‡‡  CHRISTELE PELADE§§  DOMINIQUE BABUTY
Institution:From the Department of Cardiology B, Centre Hospitalier Universitaire, Tours;, Department of Cardiology, Centre Hospitalier Universitaire Minjoz, Besançon;, Department of Cardiology, Centre Hospitalier Général, Auxerre;, Department of Cardiology, Clinique Casamance, Aubagne;, Department of Cardiology, Centre Hospitalier Emile Muller, Mulhouse;, Department of Cardiology, Centre Hospitalier Fontone, Antibes;, Department of Cardiology, Centre Hospitalier Universitaire Henri Mondor, Créteil;, Department of Cardiology, Hopital Européen Georges Pompidou, Paris;, Department of Cardiology, Clinique Pontlieue, Le Mans, and;Vitatron France, Issy les Moulineaux, France
Abstract:FAUCHIER, L., et al.: Management of Atrial Tachyarrhythmias: Benefits of Pacemaker Diagnostics. The aim of this prospective multicenter study was to assess the clinical benefits of the Selection (Vitatron) pacemaker diagnostic functions (AF 1.0) in the management of AF. Forty patients (   71 ± 9  years   of age), with documented AF and conventional pacing indications, received a Selection. The AF 1.0 function of the pacemaker was programmed to document the AF burden, onset, daily distribution, duration, premature atrial beats before onset, and mode of onset of the last 12 episodes for AF episodes exceeding 180 beats/min. By comparing patients' symptoms records, patient conventional assessment at follow-up and AF 1.0 data, the investigators evaluated the usefulness of AF 1.0 in AF management at 3- and 6-month follow-ups. Usefulness was defined as a change in arrhythmia management prompted by the disclosure of AF 1.0 data. AF recurrences were recorded in 71% of the follow-ups with symptoms reported by patients in only 16%. Thirty-nine percent of therapeutic changes based on conventional assessment were confirmed by AF 1.0 data, and in 61% of instances, the initial changes were modified by AF 1.0 data. Changes included pacing parameters in 56% of cases, AF prevention with pacing algorithms in 37%, and medical treatment in 7%. All investigators indicated that AF 1.0 was useful in all patients. The AF 1.0 diagnostic functions offered a unique documentation of AF in asymptomatic patients, and allowed therapeutic adjustments impossible otherwise. (PACE 2003; 26Pt. II]:233–238)
Keywords:atrial fibrillation                                              pacemaker                                              diagnostics functions
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号