种植体周围骨内应力分布的三维有限元分析
引用本文:王晓洁,罗教明,杨立,陈继镛,张兴栋. 种植体周围骨内应力分布的三维有限元分析[J]. 实用口腔医学杂志, 2005, 21(5): 592-596
作者姓名:王晓洁  罗教明  杨立  陈继镛  张兴栋
作者单位:成都四川大学生物材料工程研究中心,610064
基金项目:国家自然科学基金资助(批准号:30471758)
摘    要:目的针对不同类型的牙槽骨科学地选用种植体,提高种植体临床疗效,延长使用寿命。方法采用三维有限元分析方法,将圆柱状、螺纹状和台阶状种植体分别植入4类骨质结构中,对此12种情况进行应力分析。结果在同种骨质模型中,圆柱状种植体颈部周围骨内的应力集中最小;就同种形态种植体而言,较低的骨质密度不利于种植体的应力分布。结论圆柱状是一种最有利于降低颈部骨质吸收的形态结构。螺纹状种植体周围骨内应力最大值大于圆柱状,而螺纹自身非力学优势极大的拓展了该型种植体的使用范围,但螺纹尖端处的高应力区域和螺纹之间的低应力区域是影响其长期使用效果的潜在不利因素。台阶状种植体相对较适合骨质好的情况,其根部出现局部高应力区域,若应力处于骨生理承受范围之内,将有利于减少根部骨质疏松。

关 键 词:三维有限元分析  牙种植体  骨密度  应力分布
文章编号:1001-3733(2005)-05-0592-05
收稿时间:2004-12-18
修稿时间:2005-05-19

Stress distribution in bone around dental implants studied by three-dimensional finite element analysis
Wang XiaoJie;Luo JiaoMing;Yang Li;Chen JiYong;Zhang XingDong. Stress distribution in bone around dental implants studied by three-dimensional finite element analysis[J]. Journal of Practical Stomatology, 2005, 21(5): 592-596
Authors:Wang XiaoJie  Luo JiaoMing  Yang Li  Chen JiYong  Zhang XingDong
Abstract:Objective:To study the stress distribution in bone with different densities around dental implants in different shape.Methods:Dental implants in the shape of cylinder,thread and steps were implanted into the mandibular bone at the position of the first molar with 4 different densities respectively.Three-dimension finite element models were rebuilt and stress distribution in the bone around implants were investigated. Results:In the same load and bone density conditions, cylindrical implant resulted in lowest peak stress value,the same implant in mandibular bone with higher density produced lower stress peak value.Conclusion:Dental implatns in the shape of cylinder and higher density of the bone around the implant may reduce stress concentration.
Keywords:Three-dimensional finite element analysis (3D FEA)  Dental implants  Bone density  Stress distribution