首页 | 本学科首页   官方微博 | 高级检索  
     

Beckman CX9型自动控制函数变化率质量的结果
引用本文:史直心,何红梅,袁平宗,李传达,陈志远,黄作富. Beckman CX9型自动控制函数变化率质量的结果[J]. 现代检验医学杂志, 2001, 16(3): 38-39
作者姓名:史直心  何红梅  袁平宗  李传达  陈志远  黄作富
作者单位:内江市第二人民医院
摘    要:目的分析BeckmanCX9型自动生化分析仪的函数变化率质量控制(简称函变质控)效果.方法根据每日生化测定前对CX9型化学分析法的模拟数字转换(Analog-to-DigitalConversion,ADC)值的校准,选择各化学项目灵敏度(Span)和反应比率(Rate)的ADC值作统计和比较.结果各化学项目的Span值可因分析法、试剂和化学物质不同而异,但各有自身的变化规律和范围(-s);在Span的控制测定条件下,各化学项目控制误差的能力以葡萄糖(GLU)和总蛋白(TP)的测定为首,不仅浓度间的Rate值符合总体函变质控统计参数的要求(P>0.05),而且始终能保持临床室内质控(IQC)的精密度(Back-to-Back,B-B)和准确度(Range,R).结论各化学项目Span值的校准是方法学内外误差信息的综合分析结果,这有利于Span与化学反应速率的双向性相适应,达到化学反应速率的质量测定水平.

关 键 词:ADC  Span  IQC  化学反应速率  函数变化率质量控制
文章编号:1005-1716(2001)03-038-02
修稿时间:2001-02-26

The Results of Function Change Rate for Automatic Quality Control on Beckman CX9
SHI De,HE Hong-mei,YUAN Ping-zong,LI Chuan-da,CHEN Zhi-yuan,HUANG Zuo-fu. The Results of Function Change Rate for Automatic Quality Control on Beckman CX9[J]. Journal of Modern Laboratory Medicine, 2001, 16(3): 38-39
Authors:SHI De  HE Hong-mei  YUAN Ping-zong  LI Chuan-da  CHEN Zhi-yuan  HUANG Zuo-fu
Abstract:Aim To analyze the effects of function change rate for quality control on automatic biochemics analysis instrument,Beckman CX9.Methods Through checking the calibrated results of ADC(analog-to-digital conversion) of chemical analyses befor tests were done,we chose ADC of span and ADC of reaction rate to statistics and compared.Results Span of chemical analyses varie according to different analyses,chemical material and reagents,but items for testing had their own change pattern and range .The abilities of controlling errors for testing were maily in determination of glucoses (GLU) and total proteins (TP) in the controlling of tested span.The rate of consistency was not only in accordance with the demand (P>0.05) of general function alter on statistics parameter of IQC (internal quality control),but also maintained the precision (B-B)and accuracy (R) in IQC.Conclusion Calibrated span of chemical analyses are comprehesive analysis outcome from handling inside and outside error information with methodology which can make span adapt to biphase of chemical reaction rate and achieve the determination level of quality in the speed of chemical reaction.
Keywords:ADC  IQC  Span  Speed of chemical reaction  Function change rate for quality control
本文献已被 CNKI 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏