首页 | 本学科首页   官方微博 | 高级检索  
     


Pull-out strength of a polished tapered stem is improved by placing bone cement over the shoulder of the implant
Authors:Subramanian Kanthalu N  Temple Andrew J  Evans Sam  John Alun
Affiliation:Department of Trauma and Orthopaedic Surgery, University Hospital of Wales, Cardiff, UK.
Abstract:Displacement of a polished stem while attempting closed reduction of a dislocated total hip or during dislocation itself is a rare but significant complication. Our aim was to assess whether applying bone cement over the shoulder of the implant can help to prevent this. We conducted an in vitro mechanical study with tensile testing machine. We cemented 7 sawbones with a standard cementing technique and another 7 with additional cement over the shoulder of the implant. The mean pull-out force in the routine cementing technique was 2066 N (SD, 256.65), and it was 3220 N (SD, 312.22) for the group with the cement on the shoulder. There is a statistically significant difference of 1154 N. We recommend that when a polished stem is used, bone cement should be applied over the shoulder of the implant.
Keywords:polished stem   stem displacement   dislocation   cement on shoulder   pull-out force   biomechanical test
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号