基于锥形束CT体部肿瘤图像引导放疗的摆位误差分析
引用本文:吴钦宏,李高峰,王婵,李雪南,王丹,沙翔燕.基于锥形束CT体部肿瘤图像引导放疗的摆位误差分析[J].中华放射肿瘤学杂志,2012,21(1):68-71.
作者姓名:吴钦宏  李高峰  王婵  李雪南  王丹  沙翔燕
作者单位:100730 北京,卫生部北京医院放疗科
摘    要:目的 应用千伏级锥形束CT (KVCBCT)评价采用负压成型垫作体位固定和最终等中心标记法作定位的体部肿瘤图像引导放疗的摆位误差。方法 回顾分析2009—2011年的223例体部肿瘤患者资料,这些患者在配有LAP可移动式激光定位系统的飞利浦 PQS CT或飞利浦 BrillianceCT Big Bore上采用负压成型垫体位固定和最终等中心标记法定位。CT图像通过网络传输给瓦里安Eclipse治疗计划系统用来勾画靶区和设计计划。在治疗前使用瓦里安直线加速器的机载影像系统行KVCBCT扫描和配准,得出左右、上下和前后方向的摆位误差。使用SPSS 16.0软件对数据行独立样本t检验。

关 键 词:体层摄影术  X线计算机  锥形束  图像引导放疗  摆位误差  外放边界  体部肿瘤  
收稿时间:2011-04-20

Analysis of setup errors on CBCT - based image guided radiation therapy for body tumors patients
WU Qin-hong,LI Gao-feng,WANG Chan,LI Xue-nan,WANG Dan,SHA Xiang-yan.Analysis of setup errors on CBCT - based image guided radiation therapy for body tumors patients[J].Chinese Journal of Radiation Oncology,2012,21(1):68-71.
Authors:WU Qin-hong  LI Gao-feng  WANG Chan  LI Xue-nan  WANG Dan  SHA Xiang-yan
Institution:Department of Radiation Oncology, Beijing Hospital, Ministry of Health, Beijing 100730, China
Abstract:Objective To evaluate setup errors of image guided radiation therapy (IGRT) for body tumors immbilized with vacuum cushions and localized with final isocenter marked method using kilovoltage cone beam CT (KVCBCT).Methods A retrospective study has been carried out for 223 patients from March 2009 to April 2011.All patients were immobilized with vacuum cushions and localized with final isocenter marked method in Philips PQS CT or Philips Brilliance CT Big Bore scanner,which were equipped with LAP movable laser systems. The CT images were transferred to a Varian Eclipse workstation for contouring and treatment planning.Before irradiation,a KVCBCT scan was performed and image registration was done on a Varian iX linear accelerator via OBI system.Each set of setup errors in right-left ( RL),superior-inferior (SI),and anterior-posterior (AP) directions were gathered.An independent-samples t-test statistical analysis was conducted with the 758 sets of data using SPSS 16.0.Results The statistical analysis showed that setup errors from the 758 datasets were depicted a Gaussian distribution.The system errors ± random errors in RL,SI and AP were ( -0.5 ±2.8) mm,(0.0±3.0) mm and (0.4±3.4)mm,respectively.Referring to the formula for planning target volume margin calculation,M =2.5Σ + 0.7δ,the margins were calculated as 3.2,2.1 and 3.4 mm,respectively.Conclusions The margins derived from this retrospective study have confirmed the premise that the treatment plans were executed in patients with high reliability,thereby created a high sense of confidence for the clinicians.
Keywords:Tomography  X-ray computed  cone beam  Image guided radiotherapy  Setup error  Margin  Body neoplasms