Sigma metric analysis for performance of creatinine with fresh frozen serum
Authors:Fengfeng Kang  Chuanbao Zhang  Wei Wang
Institution:1. Zhejiang Center for Clinical Laboratories, Zhejiang Provincial People’s Hospital, Zhejiang, P. R. China;2. National Center for Clinical Laboratories, Beijing Hospital, Beijing, P. R. China
Abstract:Background: Six sigma provides an objective and quantitative methodology to describe the laboratory testing performance. In this study, we conducted a national trueness verification scheme with fresh frozen serum (FFS) for serum creatinine to evaluate its performance in China. Methods: Two different concentration levels of FFS, targeted with reference method, were sent to 98 laboratories in China. Imprecision and bias of the measurement procedure were calculated for each participant to further evaluate the sigma value. Quality goal index (QGI) analysis was used to investigate the reason of unacceptable performance for laboratories with σ?Results: Our study indicated that the sample with high concentration of creatinine had preferable sigma values. For the enzymatic method, 7.0% (5/71) to 45.1% (32/71) of the laboratories need to improve their measurement procedures (σ??1.2). Only 3.1–5.3% of the laboratories should improve both of the precision and trueness. Conclusions: Sigma metric analysis of the serum creatinine assays is disappointing, which was mainly due to the unacceptable analytical bias according to the QGI analysis. Further effort is needed to enhance the trueness of the creatinine measurement.
Keywords:Creatinine  six sigma  fresh frozen serum  bias  quality control