Evaluation of CT coronary artery angiography with 320-row detector CT in a high-risk population
Authors:Gang S  Min L  Li L  Guo-Ying L  Lin X  Qun J  Hua Z
Institution:Department of Medical Imaging, Jinan Military General Hospital, Jinan, Shandong Province, China. cjr.sungang@vip.163.com
Abstract:

Objectives

The aim of this article was to prospectively evaluate the accuracy and radiation dose of 320-detector row dynamic volume CT (DVCT) for the detection of coronary artery disease (CAD) in a high-risk population.

Methods

60 patients with a high risk of CAD underwent DVCT without preceding heart rate control and also underwent invasive coronary angiography (ICA), which served as the standard reference.

Results

On a per segment analysis, overall sensitivity was 95.3%, specificity was 97.6%, positive predictive value was 90.6%, negative predictive value was 98.8% and Youden index was 0.93. In both heart rate subgroups, diagnostic accuracy for the assessment of coronary artery stenosis was similar. The accuracy of the subgroup with an Agatston score ≥100 was lower than that for patients with an Agatston score <100. However, the difference between DVCT and ICA results was not significant (p=0.08). The mean estimated effective dose of CT was 12.5±9.4 mSv. In those patients with heart rates less than 70 beats per minute (bpm), the mean radiation exposure of DVCT was 5.2±0.9 mSv. The effective radiation dose was significantly lower than that of ICA (14.1±5.9 mSv) (p<0.001). When the heart rate was >70 bpm, a significantly higher dose was delivered to patients with DVCT (22.6±5.2 mSv, p<0.001) than with ICA (15.0±5.3 mSv, p<0.001).

Conclusion

DVCT reliably provides high diagnostic accuracy without heart rate/rhythm control. However, from a dosimetric point of view, it is recommended that heart rate should be controlled to <70 bpm to decrease radiation dose.The small diameter of the coronary segments, their complex three-dimensional geometry and their rapid movement throughout the cardiac cycle represent the major challenges for artefact-free coronary CT angiography (CTA). With each scanner generation, motion artefacts re-appear as a major cause of image quality degradation during coronary CTA 1-10]. Coronary CTA studies of each coronary artery with four-multidetector CT (MDCT) at a gantry rotation time of 500 ms had significantly decreased image quality with increasing mean heart rates 3]. Using 16-MDCT at a gantry rotation time of 420 ms, Hoffmann et al 2] found a significant negative correlation between overall image quality and mean heart rate. Even using 64-section CT, with its gantry rotation speed of 330 ms, elevated and irregular heart beats were found to cause relevant degradation of image quality 1,4,9,11]. Using dual-source CT (DSCT) with an increased temporal resolution of 83 ms, there was no significant correlation between mean heart rate and the overall image quality for any coronary segment or for any individual coronary artery. Nonetheless, irregular heart rates still slightly affect the image quality of non-invasive coronary angiography, even with DSCT 10,12].The 320-detector row dynamic volume CT (DVCT) is characterised by 320 slice detectors with a thickness of 0.5 mm and gantry rotation time of 350 ms. With a wide coverage of 16 cm in the z-axis, the whole heart can be covered within one cardiac cycle. Theoretically, DVCT makes it possible to scan patients with an irregular heart rate without “stair-step” artefacts. At the same time, DVCT avoids the overlapping rotations of helical CT, and the application of prospective echocardiogram (ECG) gating has become more feasible. Recent studies of DVCT have mainly been based on a low heart rate 13-17]. Few studies have investigated the diagnostic accuracy in higher heart rates and arrhythmia. Our purpose was to systematically evaluate the diagnostic accuracy and exposure dose of DVCT in a high-risk population with high and irregular heart rates.