Semiconductor real-time quality assurance dosimetry in brachytherapy
Affiliation:1. Medical Physics Unit, Ente Ospedaliero Cantonale EOC, Bellinzona, Switzerland;2. Medical Physics Unit, Fondazione IRCCS Istituto Nazionale dei Tumori, Milano, Italy;3. Centre for Medical Radiation Physics, University of Wollongong, Wollongong, Australia;4. St. George Cancer Care Centre, St George Hospital, Kogarah, Australia;5. Radiation Oncology Department, Prince of Wales Hospital, Randwick, Australia
Abstract:With the increase in complexity of brachytherapy treatments, there has been a demand for the development of sophisticated devices for delivery verification. The Centre for Medical Radiation Physics (CMRP), University of Wollongong, has demonstrated the applicability of semiconductor devices to provide cost-effective real-time quality assurance for a wide range of brachytherapy treatment modalities. Semiconductor devices have shown great promise to the future of pretreatment and in vivo quality assurance in a wide range of brachytherapy treatments, from high-dose-rate (HDR) prostate procedures to eye plaque treatments. The aim of this article is to give an insight into several semiconductor-based dosimetry instruments developed by the CMRP. Applications of these instruments are provided for breast and rectal wall in vivo dosimetry in HDR brachytherapy, urethral in vivo dosimetry in prostate low-dose-rate (LDR) brachytherapy, quality assurance of HDR brachytherapy afterloaders, HDR pretreatment plan verification, and real-time verification of LDR and HDR source dwell positions.
Keywords:Real-time QA  Brachytherapy  MOSkin  Magic Plate  BrachyView
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号