Development and characterization of an anthropomorphic breast phantom for permanent breast seed implant brachytherapy credentialing |
| |
Affiliation: | 1. Department of Physics and Astronomy, University of Calgary, Calgary, Alberta, Canada;2. Department of Medical Physics, Tom Baker Cancer Centre, Calgary, Alberta, Canada;3. Department of Oncology, Tom Baker Cancer Centre, Calgary, Alberta, Canada |
| |
Abstract: | PurposeTo develop an anthropomorphic breast phantom for use in credentialing of permanent breast seed implant brachytherapy.Methods and MaterialsA representative external contour and target volume was used as the basis of mold manufacturing for anthropomorphic breast phantom development. Both target and normal tissue were composed of gel-like materials that provide suitable computed tomography and ultrasound contrast for brachytherapy delivery. The phantoms were evaluated for consistency in construction (target location) and Hounsfield unit (computed tomography contrast). For both target and normal tissue, the speed of sound was measured and compared to the image reconstruction algorithm's expectation value. Five phantoms were imaged preimplant and postimplant to assess interphantom similarity as well as to evaluate the uncertainty in quantifying seed position.ResultsThe average Hounsfield units of the target and normal tissue gels is −146 ± 5 and 23 ± 1, respectively. The average speed of sound of the target and normal tissue gels is 1485 ± 7 m/s and 1558 ± 9 m/s, respectively, resulting in an estimated 0.4 mm uncertainty in image guidance. The registration/deformation uncertainty was determined to be 0.8 mm. The standard combined uncertainty in assessing seed position spatial accuracy, also including a 0.9 mm estimate based on literature for seed localization, is estimated to be 1.3 mm.ConclusionsThe development of the anthropomorphic breast phantom and evaluation of both the consistency as well as overall seed position uncertainty illustrates the suitability of this phantom for use in brachytherapy end-to-end delivery and implant accuracy evaluation. When evaluating a user's implant accuracy, we estimate a standard combined uncertainty of 1.3 mm. |
| |
Keywords: | Anthropomorphic breast phantom Permanent breast seed implant Brachytherapy Phantom characterization Credentialing |
本文献已被 ScienceDirect 等数据库收录! |
|