Inert gas rebreathing: the effect of haemoglobin based pulmonary shunt flow correction on the accuracy of cardiac output measurements in clinical practice
Authors:F. Trinkmann  T. Papavassiliu  F. Kraus  H. Leweling  Stefan O. Schoenberg  M. Borggrefe  J. J. Kaden  J. Saur
Affiliation:1. First Department of Medicine – Cardiology, Angiology, Pneumology, Intensive Care, Medical Faculty Mannheim, University of Heidelberg, Mannheim, Germany;2. Fourth Department of Medicine – Geriatry, Clinical Nutrition, Medical Faculty Mannheim, University of Heidelberg, Mannheim, Germany;3. Institute of Clinical Radiology and Nuclear Medicine, Medical Faculty Mannheim, University of Heidelberg, Mannheim, Germany
Abstract:Background: Cardiac output (CO) is an important cardiac parameter, however its determination is difficult in clinical routine. Non‐invasive inert gas rebreathing (IGR) measurements yielded promising results in recent studies. It directly measures pulmonary blood flow (PBF) which equals CO in absence of significant pulmonary shunt flow (QS). A reliable shunt correction requiring the haemoglobin concentration (cHb) as only value to be entered manually has been implemented. Therefore, the aim of the study was to evaluate the effect of various approaches to QS correction on the accuracy of IGR. Methods: Cardiac output determined by cardiac magnetic resonance imaging (CMR) served as reference values. The data was analysed in four groups: PBF without correcting for QS (group A), shunt correction using the patients’ individual cHb values (group B), a fixed standard cHb of 14·0 g dl?1 (group C) and a gender‐adapted standard cHb for male (15·0 g dl?1) and female (13·5 g dl?1) probands each (group D). Results: 147 patients were analysed. Mean COCMR was 5·2 ± 1·4 l min?1, mean COIGR was 4·8 ± 1·3 l min?1 in group A, 5·1 ± 1·3 in group B, 5·1 ± 1·3 l min?1 in group C and 5·1 ± 1·4 l min?1 in group D. The accuracy in group A (mean bias 0·5 ± 1·1 l min?1) was significantly lower as compared to groups B, C and D (0·1 ± 1·1 l min?1; P<0·01). Conclusion: IGR allows a reliable non‐invasive determination of CO. Since PBF significantly increased the measurement bias, shunt correction should always be applied. A fixed cHb of 14·0 g dl?1 can be used for both genders if the exact cHb value is not known. Nevertheless, the individual value should be used if any possible.
Keywords:cardiac output  haemoglobin –   accuracy  inert gas rebreathing  shunt
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号