An option for optimising the radiographic technique for horizontal beam lateral (HBL) hip radiography when using digital X-ray equipment
Institution:1. Department of Surgery, Reinier de Graaf Gasthuis, Reinier de Graafweg 5, 2625 AD Delft, the Netherlands;2. Trauma Research Unit Department of Surgery, Erasmus MC, University Medical Center Rotterdam, Doctor Molewaterplein 40, 3015 GD Rotterdam, the Netherlands;3. Department of Medical Physics, Reinier de Graaf Gasthuis, Reinier de Graafweg 5, 2625 AD Delft, the Netherlands;4. Faculty of 3ME, Delft University of Technology
Abstract:AimTo investigate the optimum technique for the horizontal beam lateral (HBL) hip projection considering image quality and radiation dose.MethodsUsing digital radiography equipment an anthropomorphic phantom was positioned for a HBL projection of the hip. Radiographic exposures were undertaken across a range of acquisition parameters (tube potentials, source to image distances, object to detector distances, with and without an anti-scatter radiation grid/additional copper filtration). Each acquisition combination was imaged three times and the dose area product (DAP) and post-AEC mAs recorded. 168 images were acquired. A single observer evaluated five anatomical areas on all images using a two-alternative force choice technique. The reference image was selected based on the current locally accepted technique. 50 images out of the original 168 were independently assessed by a further four observers to ensure reliability of the results.ResultsImage quality, when comparing all the images to the reference, was improved on in two cases; however the radiation dose had increased. 18 images had equal image quality with some having an 80% reduction in the DAP. In terms of the diagnostic acceptability, 51 were considered acceptable with a lower radiation dose.ConclusionBy optimising acquisition factors for the HBL hip projection the radiation dose to the patient can be reduced. Based on the findings the factors proposed for HBL hip projections are 90 kVp, 135 cm SID, 45 cm ODD, grid and 0.1 mm copper filtration.
Keywords:Image quality  Radiation dose  Anthropomorphic phantom  Hip radiography  Dose optimisation  Digital radiography
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号