Increased diffusion sensitivity with hyperechos.
Authors:Lawrence R Frank  Eric C Wong  Thomas T Liu  Richard B Buxton
Institution:UCSD Center for Functional MR1, San Diego, California 92093, USA. lfrank@ucsd.edu
Abstract:It is shown that the introduction of a 180 degrees refocusing pulse into a standard diffusion weighted stimulated echo sequence is equivalent to the simplest hyperecho sequence with identical diffusion weighting but equal or greater signal-to-noise (SNR) and thus equal or greater diffusion contrast. For high b-value imaging, the hyperecho sequence thus possesses the high diffusion contrast in the presence of small T(1)/T(2) ratios characteristic of stimulated echo sequences but with less than the 50% loss in SNR that is associated with the stimulated echo. For low b-value imaging, the hyperecho signal converges to that of the standard spin echo. The advantages of the two-pulse diffusion weighted hyperecho sequence are demonstrated theoretically. Experimental results are shown in the application to high angular resolution diffusion encoding (HARD) in normal human brain.
Keywords:diffusion  hyperechoes  high angular resolution diffusion  diffusion anisotropy