我院脉动真空灭菌器灭菌质量的监测
引用本文:戴李中,郑柏湘. 我院脉动真空灭菌器灭菌质量的监测[J]. 医疗卫生装备, 2006, 27(7): 64-64
作者姓名:戴李中  郑柏湘
作者单位:绍兴市人民医院,浙江省,绍兴市,312000
摘    要:目的:为了解脉动真空灭菌器的灭菌质量,确保灭菌效果.方法:采用物理、化学、生物方法,对其灭菌质量进行监测.结果:经监测4 576锅次,温度监测合格率为99.6%,包内化学指示卡监测合格率达到99.8%,B-D试验监测合格率99.9%,生物指示剂监测合格率100%.结论:该院脉动真空灭菌器运行状况良好,物理、化学、生物监测合格率达到99%以上.

关 键 词:脉动真空灭菌器  灭菌质量  监测
文章编号:1003-8868(2006)07-0064-01
收稿时间:2005-11-21
修稿时间:2006-06-22

Monitoring of sterilization quality of pulsating vacuum sterilizer in our hospital
DAI Li-zhong,ZHENG Bai-xiang. Monitoring of sterilization quality of pulsating vacuum sterilizer in our hospital[J]. Chinese Medical Equipment Journal, 2006, 27(7): 64-64
Authors:DAI Li-zhong  ZHENG Bai-xiang
Abstract:Objective To find out sterilization quality of pulsating vacuum sterilizer for better sterilization effect. Methods The sterilization quality was monitored by physical, chemical and biological methods. Results 4,576 subjects were involved in the test. Such data were obtained as the qualification rate for temperature, the one by chemical indicator card, the one by B-D test and the one by biological indicator, which were 99.6%, 99.8%, 99.9% and 100% respectively. Conclusion Above data proves that the sterilizer is in normal condition.
Keywords:pulsating vacuum sterilizer   sterilizing quality   monitoring