Activated Pyrolytic Carbon Tip Pacing Leads: An Alternative to Steroid-Eluting Pacing Leads?
Authors:MICHELE MASINI  MAURO LAZZARI  ROBERTO LORENZONI  ANNA MARIA DOMICELLI  AGNESE MICHELETTI  RENZO DIANDA  GIUSEPPE MASINI
Affiliation:Cardiology Department, "Campo di Marte" Hospital, Lucca, Italy
Abstract:Pyrolytic carbon technology is known for its excellent mechanical properties and electrical conductivity; it is particularly biocompatible and does not require high production costs. The Sorin S100/4 lead is a ventricular passive fixation lead with a hemispherical electrode surface area of 4 mm2; the stimulating tip is made of a graphite core coated by a thin activated pyrolytic carbon layer. We evaluated the acute and medium-term performance of the unipolar version of this lead in 65 patients. At implantation, pacing threshold (at 0.5 ms) was 0.26 ± 0.08 V; pacing impedance (at 5 V and 0.5 ms) was 537 ± 94 Ω and Ft wave amplitude was 15.0 ± 5.5 mV. No lead related complications (dislodgment, perforation, exit block, etc.) occurred in any patient; one patient presented with a wire fracture after 26 months, due to subclavian crush syndrome. Follow-up procedures were performed at 1 week, and 2-, 10-, 18-, and 30-months postimplant. Since the leads were connected to pacemakers from different manufacturers, either voltage or duration thresholds were measured, In approximately two thirds of the patients, with an output of 2.5 V, a mean duration threshold of 0.16 ± 0.13 ms at 1 week, 0.12 ± 0.08 ms at 2 months, 0.11 ± 0.06 ms at 10 months, 0.09 ± 0.06 ms at 18 months, and 0.07 ± 0.03 ms at 30 months, was measured. In the remaining one-third of the patients, a comparable voltage threshold trend was measured. The mean pacing impedance showed a transient drop at 1 week, and then increased to a plateau of about 600 fl reached after 10 months. No sensing defect occurred in any patient. Our data show good acute and intermediate-term results of the S100/4 lead; the early rise in threshold was remarkably blunted. Activated pyrolytic carbon tip leads might therefore be considered as a possible, inexpensive alternative to steroideluting leads.
Keywords:pacemaker leads    steroid elution    pacing threshold    pacing impedance    follow-up
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号