股骨头人工关节对盆腔放疗剂量分布的影响
引用本文:任江平,张松方,朱其宝,郭建新,查元梓. 股骨头人工关节对盆腔放疗剂量分布的影响[J]. 中华放射肿瘤学杂志, 2011, 20(4). DOI: 10.3760/cma.j.issn.1004-4221.2011.04.025
作者姓名:任江平  张松方  朱其宝  郭建新  查元梓
作者单位:1. 鄞州人民医院放化疗中心,宁波,315040
2. 上海交通大学医学院附属新华医院肿瘤科
摘    要:目的 测量钴、镉、钼合金股骨头人工关节在放疗过程中次级射线引起的散射效应,以及其作为高原子序数材料对剂量分布的影响.方法 测量6、10 MV X射线入射方向距关节0.5、1.0、2.0 cm处剂量,出射方向距关节3.0、5.0、7.0 cm处剂量及距关节头部5.0、10.0 cm的剂量曲线.结果 6、10 MV X射线入射方向距关节0.5、1.0、2.0 cm处剂量偏差在0%~5%之间,随能量增加背散射效应更加显著.射线出射方向距关节3.0、5.0、7.0 cm处剂量偏差为21.6%~30.8%,射野和深度相同时随能量增加剂量偏差减少,同一能量和深度下剂量偏差随射野增加而减少.透射剂量曲线显示关节头部对剂量衰减作用极为显著.结论 在进行盆腔放疗时尽量避免射野穿过关节,或者至少降低穿越关节的射野权重.
Abstract:
Objective To study the scattering effect of Co-Cr-Mo hip prosthesis which was high Z material for patients undergoing pelvic irradiation.Methods The hip prosthesis was set in water phantom (30 cm×30 cm×30 cm), determing points were chosen on the entrance side of both 6 MV and 10 MV beams at the distance of 0.5 cm, 1.0 cm, 2.0 cm to the hip prosthesis, and also on the exit side of both 6 MV and 10 MV beams at the distance of 3.0 cm, 5.0 cm, 7.0 cm to the hip prostheses.Dose behind the hip prosthesis at depths of 5.0 cm and 10.0 cm for 6 MV and 10 MV beams are also measured.ResultsThe dose deviation on the beams′ entrance side is between 0 to 5.0%, the backscatter effect was more obviously with the higher energy beam.The dose deviation on the beams′ exit side was between 21.6%-30.8%.With the same field size and depth, dose deviation becomes smaller when the beam energy was higher;while with the same energy and depth, dose deviation becomes smaller when the field size was bigger.Dose profiles behind the head of the hip prosthesis indicate obvious attenuation of the beam.Conclusions Beam arrangements that avoid the prosthesis should be considered first or we should at least reduce the weight of the beam that pass through the prosthesis.

关 键 词:髋假体,金属  X线照射  剂量学

Impact of hip prosthesis on dose distribution of pelvic radiotherapy
REN Jiang-ping,ZHANG Song-fang,ZHU Qi-bao,GUO Jian-xin,ZHA Yuan-zi. Impact of hip prosthesis on dose distribution of pelvic radiotherapy[J]. Chinese Journal of Radiation Oncology, 2011, 20(4). DOI: 10.3760/cma.j.issn.1004-4221.2011.04.025
Authors:REN Jiang-ping  ZHANG Song-fang  ZHU Qi-bao  GUO Jian-xin  ZHA Yuan-zi
Abstract:Objective To study the scattering effect of Co-Cr-Mo hip prosthesis which was high Z material for patients undergoing pelvic irradiation.Methods The hip prosthesis was set in water phantom (30 cm×30 cm×30 cm), determing points were chosen on the entrance side of both 6 MV and 10 MV beams at the distance of 0.5 cm, 1.0 cm, 2.0 cm to the hip prosthesis, and also on the exit side of both 6 MV and 10 MV beams at the distance of 3.0 cm, 5.0 cm, 7.0 cm to the hip prostheses.Dose behind the hip prosthesis at depths of 5.0 cm and 10.0 cm for 6 MV and 10 MV beams are also measured.ResultsThe dose deviation on the beams′ entrance side is between 0 to 5.0%, the backscatter effect was more obviously with the higher energy beam.The dose deviation on the beams′ exit side was between 21.6%-30.8%.With the same field size and depth, dose deviation becomes smaller when the beam energy was higher;while with the same energy and depth, dose deviation becomes smaller when the field size was bigger.Dose profiles behind the head of the hip prosthesis indicate obvious attenuation of the beam.Conclusions Beam arrangements that avoid the prosthesis should be considered first or we should at least reduce the weight of the beam that pass through the prosthesis.
Keywords:Hip prosthesis,material  X-ray irradiation  Dosimetry