Influence of tube potential on quantitative coronary plaque analyses by low radiation dose computed tomography: a phantom study
Authors:Chunhong Wang  Yuliang Liao  Haibin Chen  Xin Zhen  Jianhong Li  Yikai Xu  Linghong Zhou
Institution:1.Department of Radiology,Xinyang Central Hospital,Xinyang,China;2.Department of Biomedical Engineering,Southern Medical University,Guangzhou,China;3.Department of Medical Imaging Center, Nanfang Hospital,Southern Medical University,Guangzhou,China
Abstract:Previous studies have shown that employing the low dose computed tomography (CT) technique based on low tube potential reduces the radiation dose required for the coronary artery examination protocol. However, low tube potential may adversely influence the CT number of plaque composition. Therefore, we aimed to determine whether quantitative atherosclerotic plaque analysis by a multi-slice, low radiation dose CT protocol using 80 kilovolts (kV) yields results comparable to those of the standard 120 kV protocol. Artificial plaque samples (n?=?17) composed of three kinds of plaque were scanned at 120 and 80 kV. Relative low-density and medium-density plaque component volumes obtained by three protocols (80 kV, 60 Hounsfield units HU] threshold; 120 kV, 60 HU threshold; and 80 kV, 82 HU threshold) were compared. Using the 60 HU threshold, relative volume of the low-density plaque component obtained at 80 kV was lower than that obtained at 120 kV (27?±?3% vs. 51?±?5%, P?<?0.001), whereas relative volume of the medium-density plaque component obtained at 80 kV was higher than that obtained at 120 kV (73?±?3% vs. 48?±?5%, P?<?0.001). By contrast, no significant difference in relative volume obtained at 80 kV (82 HU threshold) versus 120 kV (60 HU threshold) was observed for either low-density (52?±?5% vs. 51?±?5%) or medium-density (48?±?5% vs. 48?±?5%) plaque component. Low tube potential may affect the accuracy of quantitative atherosclerotic plaque analysis. For our phantom test, 82 HU was the optimal threshold for scanning at 80 kV.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号