首页 | 本学科首页   官方微博 | 高级检索  
     

适配子型压电石英晶体传感器探针固定方法的研究
引用本文:齐永志,府伟灵,姚春艳. 适配子型压电石英晶体传感器探针固定方法的研究[J]. 第三军医大学学报, 2007, 29(5): 399-401
作者姓名:齐永志  府伟灵  姚春艳
作者单位:第三军医大学西南医院检验科,重庆,400038;第三军医大学西南医院检验科,重庆,400038;第三军医大学西南医院检验科,重庆,400038
摘    要:目的 比较金膜表面两种探针固定法的优劣,选择适配子型压电传感器探针固定的最佳方法.方法 用巯基固定法及生物素-亲和素固定法将针对人IgE的适配子探针固定在压电传感器的金膜表面,对不同浓度IgE引起的频率变化进行检测.结果 与巯基法相比,生物素-亲和素法固定探针压电传感器频率下降更为明显,检出限低,线性范围宽,0.1~2.5 mg/L浓度范围内IgE浓度与频率变化呈明显的线性相关,相关系数0.994 5.结论 在压电基因传感器探针固定中效果良好的巯基法在适配子探针固定中并不适用,生物素-亲和素法在适配子压电传感器探针固定中有明显优势.

关 键 词:石英晶体微天平  传感器  适配子  探针
文章编号:1000-5404(2007)05-0399-03
修稿时间:2006-07-27

Immobilization methods for probes of aptamer-based piezoelectric quartz crystal sensor
QI Yong-zhi,FU Wei-ling,YAO Chun-yan. Immobilization methods for probes of aptamer-based piezoelectric quartz crystal sensor[J]. Acta Academiae Medicinae Militaris Tertiae, 2007, 29(5): 399-401
Authors:QI Yong-zhi  FU Wei-ling  YAO Chun-yan
Affiliation:Department of Clinical Laboratories, Southwest Hospital, Third Military Medical University, Chongqing 400038, China
Abstract:Objective To compare 2 methods to immobilize probes on the surface of gold membrane and select the better one for aptamer immobilization of aptamer-based piezoelectric quartz crystal sensor. Methods Anti-IgE aptamers were immobilized on the gold surface of piezoelectric quartz crystal sensor with thiol method or biotin-avidin method. The changes of frequency were compared in reactions with different concentrations of IgE. Results Biotin-avidin method obtained higher frequency changes in the reactions of IgE, lower limit of detection and wider liner range than thiol method. For biotin-avidin method, there was significant linear correlation between frequency changes and the concentration of IgE with the correlation coefficient of 0.994 5. Conclusion The thiol method which is proved effective for probe immobilization on gene sensor is unfit for aptamer immobilization. Biotin-avidin method has a better ability to immobilize probe on the surfaces of aptamer-based piezoelectric quartz crystal sensor.
Keywords:quartz crystal microbalance  sensor  aptamer  probe
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号