Biomechanical analysis of the FlatWire Figure 8 sternal fixation device
Authors:Ryan M Wilson  Paul A Ghareeb  W Thomas McClellan  Ashley N Boustany
Institution:1.University of Kentucky College of Medicine, Department of Plastic Surgery, Lexington, Kentucky;;2.Emory University School of Medicine Department of Surgery, Division of Plastic Surgery, Atlanta, Georgia;;3.West Virginia University School of Medicine, Department of Surgery, Division of Plastic Surgery, Morgantown, West Virginia;;4.University of Kentucky College of Medicine, Department of Plastic Surgery, Lexington, Kentucky, USA
Abstract:

OBJECTIVE:

To determine whether the FlatWire Figure 8 sternal fixation device (Penn United, USA) is mechanically superior to the current standard in sternotomy closure.

DESCRIPTION:

Unstable sternal closure using traditional steel-wire cerclage can increase postoperative pain, bony cut-through and wound dehiscence. The authors present the Figure 8 sternal fixation device to minimize these complications. Biomechanical properties of the device were compared with conventional steel wire sternal repair.

EVALUATION:

Using two constructs of both FlatWire and steel wire, pull-to-failure, Hertzian contact and cut-through were compared. Samples were tested to 500,000 cycles or failure. Cyclic comparisons were performed using log-rank t tests and Student’s t tests for cut-through analysis. FlatWires were found to have superior biomechanical properties in all categories tested.

CONCLUSION:

The FlatWire provides superior biomechanical properties compared with conventional steel wire, which may lead to reduced sternal wound complications.
Keywords:Figure 8  FlatWire  Sternal fixation  Sternal wound  Sternotomy