Guglielmi detachable coil extraction: suction generated by pusher-wire withdrawal after coil detachment within an intracranial aneurysm |
| |
Authors: | Ng Perry P Halbach Van V Caragine Louis P Dowd Christopher F Higashida Randall T |
| |
Affiliation: | Division of Interventional Neuroradiology, University of California at San Francisco Medical Center, San Francisco, USA. |
| |
Abstract: | We present a case of Guglielmi detachable coil extraction into a microcatheter after coil detachment during embolization of a superior hypophyseal artery aneurysm; extraction was a result of the suction generated during pusher-wire withdrawal. Experimental simulations using many coil and microcatheter combinations were used to identify factors that contribute to the likelihood of this phenomenon. |
| |