Ferromagnetism and MR imaging: safety of carotid vascular clamps
Authors:G P Teitelbaum  M C Lin  A T Watanabe  J F Norfray  T I Young  W G Bradley
Institution:Department of Radiology, Georgetown University Hospital, Washington, DC 20007.
Abstract:Metallic extracranial carotid vascular clamps of the Selverstone, Crutchfield, Poppen-Blaylock, Salibi, Kindt, and tantalum varieties have been placed for treatment of large, giant, or inoperable intracranial aneurysms. To ascertain what adverse effect, if any, MR imaging would have on these clamps, magnetic deflection at 1.5 T was measured for various carotid clamps. Marked magnetic deflection (and torque) was displayed by stainless steel Poppen-Blaylock clamps. Relatively mild magnetic deflection was displayed by the stainless steel Selverstone, Salibi, Crutchfield, and Kindt clamps. Three patients with previously placed carotid clamps (two Selverstone, one Salibi) and one patient with a nonferromagnetic tantalum carotid clip had cranial or cervical MR studies at field strengths ranging from 0.35 to 0.60 T. No patient experienced any discomfort or neurologic sequelae as a result of MR imaging. Although the ferromagnetic clamps created severe "black-hole" artifacts and image distortion within the cervical and facial regions, no significant image degradation was apparent during spin-echo imaging of the brain. The tantalum clip created a far smaller MR artifact than did ferromagnetic clamps and allowed effective spin-echo and gradient-echo imaging in the cervical region. Our findings indicate that most patients with carotid vascular clamps (and nonferromagnetic clips) can probably be imaged safely with MR.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号