首页 | 本学科首页   官方微博 | 高级检索  
     


Adverse reactions to an absorbable shoulder fixation device
Authors:Edwards D J  Hoy G  Saies A D  Hayes M G
Affiliation:From Sportsmed SA, Adelaide, Australia.
Abstract:Bioabsorbable implants have been advocated for use in repair of a torn glenoid labrum to correct shoulder instability. These implants have been thought to be safe and to have little toxicity, antigenicity, or adverse side effects. However, similar implants have been shown to cause nonspecific granulomatous reactions when implanted into bone or soft tissues in studies in both animals and human beings. This article reports the adverse effects of these implants in six shoulders of five patients who had repair of a damaged glenoid labrum. All patients reported increasing pain and loss of shoulder motion after insertion of these devices. All required a further arthroscopic lavage and debridement to reduce the inlracapsular synovitis. Granulomatous reactions were identified histologically in all cases. No long-term side effects were encountered, and all patients recovered satisfactorily.
Keywords:
本文献已被 PubMed 等数据库收录!