首页 | 本学科首页   官方微博 | 高级检索  
     


Relative stiffness, transverse displacement and dynamization in comparable external fixators
Authors:T.N. Gardner BSc   DPhil  M. Evans
Affiliation:

Oxford Orthopaedic Engineering Centre, Nuffield Orthopaedic Centre, Oxford, UK

Abstract:Several unilateral external fixators were laboratory tested under applied axial, bending, and torsional loads. In certain cases the shear movement at the fracture site was substantial, up to 4 mm, and in all cases it was as significant as axial displacement. Also, under average full weight bearing, all the fixator/bone screw frameworks will allow too much axial movement at the fracture. This could inhibit healing and cause peak bone and screw stresses to approach yield.
Keywords:Author Keywords: External fixators   fracture stiffness   axial   torsion   bending   displacement   bone strain   bone stress   screw stress   bone healing
本文献已被 ScienceDirect 等数据库收录!