Relative stiffness, transverse displacement and dynamization in comparable external fixators |
| |
Authors: | T.N. Gardner BSc DPhil M. Evans |
| |
Affiliation: | Oxford Orthopaedic Engineering Centre, Nuffield Orthopaedic Centre, Oxford, UK |
| |
Abstract: | Several unilateral external fixators were laboratory tested under applied axial, bending, and torsional loads. In certain cases the shear movement at the fracture site was substantial, up to 4 mm, and in all cases it was as significant as axial displacement. Also, under average full weight bearing, all the fixator/bone screw frameworks will allow too much axial movement at the fracture. This could inhibit healing and cause peak bone and screw stresses to approach yield. |
| |
Keywords: | Author Keywords: External fixators fracture stiffness axial torsion bending displacement bone strain bone stress screw stress bone healing |
本文献已被 ScienceDirect 等数据库收录! |
|