C OMPARISON OF C APILLARY AND V ENOUS G LUCOSE M EASUREMENTS IN H EALTHY V OLUNTEERS
Authors:Deborah L. Funk  Lisa Chan  Nanci Lutz  Vincent P. Verdile
Affiliation:1. Department of Emergency Medicine, Albany Medical College (DLF, VPV), Albany, New York;2. the Department of Emergency Medicine, University of Arizona College of Medicine (LC), Tucson, Arizona;3. and the Town of Guilderland Emergency Medical Services (NL), Guilderland, New York.
Abstract:Background. Out-of-hospital (OOH) emergency personnel measure serum glucose in order to determine the need for dextrose therapy. Most devices that measure serum glucose are designed to use capillary blood obtained from a finger puncture. However, OOH emergency personnel often use venous blood obtained during intravenous line (IV) placement to determine serum glucose. Objective. To compare capillary and venous glucose measurements. Methods. This prospective study used healthy, non-fasting volunteers. Simultaneous venous and capillary blood samples were obtained from each subject. Glucose levels were measured using a glucometer designed for capillary samples. The capillary and venous measurements were compared using a Pearson correlation coefficient. Power analysis revealed that the study had the ability to detect a difference of 15?mg/dL. Results. Ninety-seven volunteers (56 males, 41 females) with a mean age of 37 ± 11.9 years were enrolled. The mean capillary and venous glucose values were 104.5 ± 20.7?mg/dL and 109.7 ± 22.4?mg/dL, respectively. The Pearson correlation coefficient was 0.24. Conclusions. The correlation between venous and capillary blood glucose measurements is relatively poor in this group of healthy volunteers. Further research must be conducted on patients at risk for abnormal blood glucose.
Keywords:glucose  capillary  venous  prehospital.
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号