1. Comprehensive Health, Inc., San Francisco, CA, USA;2. Effects Technology, Inc., Santa Barbara, CA, U.S.A.
Abstract:
Temperature studies of microtitration plates demonstrate that the use of a common bacteriology incubator for heating the plates can cause a phase lag of over 30 min for the fluid in the wells to reach 37°C from ambient temperature, and that a temperature gradient of as much as 1.6°C can exist between the peripheral and center wells. This gradient is a cause of the ‘rim’ or edge effect noted in enzyme immunoassay using microtitration plates. The problem is corrected by the use of a specially designed forced air microtitration plate incubator.