Performance of the ARX-derived auditory evoked potential index as an indicator of anesthetic depth: a comparison with bispectral index and hemodynamic measures during propofol administration |
| |
Authors: | Struys Michel M R F Jensen Erik Weber Smith Warren Smith N Ty Rampil Ira Dumortier Frank J E Mestach Christel Mortier Eric P |
| |
Affiliation: | Department of Anesthesia, University Hospital of Gent, Gent, Belgium. Michel.Struys@rug.ac.be |
| |
Abstract: | BACKGROUND: Autoregressive modeling with exogenous input of middle-latency auditory evoked potential (A-Line autoregressive index [AAI]) has been proposed for monitoring anesthetic depth. The aim of the current study was to compare the accuracy of this new index with the Bispectral Index (BIS), predicted effect-site concentration of propofol, and hemodynamic measures. METHODS: Twenty female patients scheduled for ambulatory gynecologic surgery received effect compartment controlled infusion of propofol. Target effect-site concentration was started at 1.5 microg/ml and increased every 4 min by 0.5 microg/ml. At every step, sedation level was compared with monitoring values using different clinical scoring systems and reaction to noxious stimulus. RESULTS: Bispectral Index, AAI, and predicted propofol effect-site concentration were accurate indicators for the level of sedation and loss of consciousness. Hemodynamic variables were poor indicators of the hypnotic-anesthetic status of the patient. BIS correlated best with propofol effect-site concentration, followed by AAI. Hemodynamic measurements did not correlate well. No indicators predicted reaction to noxious stimulus. Poststimulus, BIS and AAI showed an increase as a result of arousal. This reaction occurred more rapidly with the AAI than with BIS. CONCLUSION: Bispectral Index, AAI, and predicted propofol effect-site concentration revealed information on the level of sedation and loss of consciousness but did not predict response to noxious stimulus. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|