Power injection of peripherally inserted central catheters
Institution:1. Kannur University, Kannur, India;2. Government Victoria College, Palakkad, India
Abstract:PURPOSETo study the tolerance of peripherally inserted central catheters (PICCs) of varying sizes and materials to power injection of radiographic contrast agents.MATERIALS AND METHODSEight different models of silicone and five different models of polyurethane single-lumen PICCs were injected with increasing rates of iothalamate 60% with use of a power injector. Tolerated and bursting rates and pressures were recorded.RESULTSThere was a wide range of tolerated rates and pressures, depending on the inner and outer diameters of the catheters and on the catheter material. Silicone PICCs tolerated rates between 0.4 and 7.0 mL/sec and polyurethane PICCs tolerated rates between 0.6 and 10.2 mL/sec, depending on the specific catheter. The 5 Fr silicone PICCs and the 4 Fr and 5Fr polyurethane PICCs tested all tolerated rates greater than 4 mL/sec. Silicone catheters tolerated pressures between 107 and 184 psi, and polyurethane catheters tolerated pressures between l60 and 314 psi.CONCLUSIONSLarger single-lumen silicone and polyurethane PICCs may be suitable for power injection of contrast agents.
Keywords: