Performance evaluation of dual-head coincidence gamma camera for positron imaging
Authors:Kaneta Tomohiro  Yamaguchi Keiichirou  Kurihara Hideyuki  Abe Yoetsu  Yamazaki Tetsuro  Hakamaduka Takashi  Sakayauchi Toru  Takai Yoshihiro  Maruoka Shin  Kinomura Shigeo  Fukuda Hiroshi  Itoh Masatoshi  Takahashi Shoki  Yamada Shogo
Affiliation:Department of Radiology, Tohoku University.
Abstract:AIM AND METHODS: We assessed the performance evaluation of a dual-head coincidence gamma camera (Millennium VG, General Electric, Milwaukee, USA) as a positron emission tomography (PET) scanner based on the measurement indicator for performance evaluation of PET device. RESULTS: The axial spatial resolution was 4.98 mm FWHM at the center of the field of view. The true high count-rate characteristic curve was relatively straight within the range 0-0.2 microCi/ml. The relative recovery coefficient was 0.1 with a diameter of 10 mm and 0.39 with a diameter of 20 mm. CONCLUSION: In this study, we had to make a some change in the protocols described in the measurement indicator for performance evaluation of PET devices.