Reduction of transient signal oscillations in true-FISP using a linear flip angle series magnetization preparation.
Authors:Vibhas S Deshpande  Yiu-Cho Chung  Qiang Zhang  Steven M Shea  Debiao Li
Institution:Department of Radiology, Northwestern University, Chicago, Illinois, USA.
Abstract:An electrocardiogram (ECG)-triggered, magnetization-prepared, segmented, 3D true fast imaging with steady-state precession (true-FISP) sequence with fat saturation was recently proposed for coronary artery imaging. A magnetization preparation scheme consisting of an alpha/2 radiofrequency (RF) pulse followed by 20 constant flip angle dummy RF cycles was used to reduce signal oscillations in the approach to steady state. However, if large resonance offsets on the order of 70-100 Hz are present, significant magnetization oscillations will still occur during data acquisition, which will result in image ghosting and blurring. The goal of this work was to validate that a linear flip angle (LFA) series can be used during magnetization preparation to reduce these image artifacts. Computer simulations, phantom studies, and coronary artery imaging in healthy volunteers were performed to compare this magnetization preparation scheme with that of an alpha/2 pulse followed by constant flip angle dummy RF cycles. The results demonstrated substantial reduction in the apparent image artifacts when using linearly increasing flip angles during magnetization preparation.
Keywords:magnetic resonance imaging (MRI)  fast imaging  steady‐state free precession (SSFP)  MR angiography  coronary arteries  image artifacts