Computer-assisted quantitative assessment of power Doppler US: effects of microbubble contrast agent in the differentiation of breast tumors |
| |
Authors: | Kettenbach Joachim Helbich Thomas H Huber Sabine Zuna Ivan Dock Wolfgang |
| |
Institution: | Department of Radiology, University Hospital of Vienna, Waehringer Guertel 18-20, A-1090 Vienna, Austria. joachim.kettenbach@univie.ac.at |
| |
Abstract: | RATIONALE AND OBJECTIVES: To objectively quantify the effects of a microbubble contrast agent to differentiate breast tumors with power doppler ultrasound and to compare these results with color doppler ultrasound (CD US). METHODS: In 47 patients a microbubble contrast agent was injected intravenously. Computer-assisted quantitative assessment of the color pixel density was performed to evaluate the increase in Doppler signals. Results were compared to previously published results of a color Doppler ultrasound study. RESULTS: Peak color pixel density at contrast-enhanced power Doppler ultrasound was higher for carcinomas than for benign tumors (P < 0.03). Time to peak enhancement was shorter in carcinomas than in benign tumors (P < 0.01). For both parameters, diagnostic accuracy of power Doppler ultrasound was 69 and 78%, and for color Doppler ultrasound 62 and 76%, respectively. CONCLUSIONS: Quantitative assessment of contrast-enhanced power Doppler ultrasound showed significant differences in malignant and benign breast tumors. Diagnostic accuracy of contrast-enhanced power Doppler ultrasound was higher compared to color Doppler ultrasound. |
| |