首页 | 本学科首页   官方微博 | 高级检索  
检索        

国产全自动血培养系统及配套血培养瓶性能评估
引用本文:张 戈,张小江,胡继红,杨 青,徐志鹏,范 欣,朱华栋,马小军,隆 云,徐英春.国产全自动血培养系统及配套血培养瓶性能评估[J].现代检验医学杂志,2018,0(6):132-135+139.
作者姓名:张 戈  张小江  胡继红  杨 青  徐志鹏  范 欣  朱华栋  马小军  隆 云  徐英春
作者单位:1.中国医学科学院北京协和医院,侵袭性真菌病机制研究与精准诊断北京市重点实验室,北京 100730; 2.卫生部北京医院,北京 100730; 3.浙江大学医学院附属第一医院,杭州 310003
摘    要:目的 通过与美国BD BACTEC FX200全自动血培养系统的比较,对国产LABSTAR全自动血培养系统进行性能评估。方法 两种血培养系统分别使用标准菌株模拟样本和临床样本,分别对需养瓶和厌氧瓶进行评估。将标准菌株和疑似血流感染的临床样本按要求分别注入两种产品配套血培养瓶内进行培养,对比评估国产LABSTAR与对照产品结果的一致率,及阳性结果报警的时间差异。结果模拟样本评估结果:需氧瓶组评估23例模拟样本,两个血培养系统全部报阳,阳性报警时间无统计学差异。厌氧瓶组评估14例模拟样本,同时报阳的11例报警有统计学差异。临床样本评估结果:需氧瓶组531例,25例报阳结果相同,报警时间有统计学差异,培养结果总体一致率为96.4%; 厌氧瓶组361例,8例报阳结果相同,报警时间有统计学差异,培养结果总体一致率为95.8%。结论 国产LABSTAR全自动血培养系统与对照美国BD BACTEC FX200全自动血培养系统培养结果一致率高,但阳性结果报警时间偏长。

关 键 词:血流感染  血培养  病原菌

Performance Evaluation of Domestic LABSTAR120 Automated Blood Culture System and Its Blood Culture Bottles
ZHANG Ge,ZHANG Xiao-jiang,HU Ji-hong,YANGQing,XU Zhi-peng,FAN Xin,ZHU Hua-dong,MA Xiao-jun,LONG Yun,XU Ying-chun.Performance Evaluation of Domestic LABSTAR120 Automated Blood Culture System and Its Blood Culture Bottles[J].Journal of Modern Laboratory Medicine,2018,0(6):132-135+139.
Authors:ZHANG Ge  ZHANG Xiao-jiang  HU Ji-hong  YANGQing  XU Zhi-peng  FAN Xin  ZHU Hua-dong  MA Xiao-jun  LONG Yun  XU Ying-chun
Institution:1.Mechanism Research and Accurate Diagnosis of Invesive Fungal Disease in Beijing Key Laboratory,Peking Union Medical College Hospital,Chinese Academy of Medical Sciences,Beijing 100730,China; 2.Ministry of Health Beijing Hospital,Beijing 100730,China; 3.the First Affiliated Hospital,Medical College of Zhejiang University,Hangzhou 310003,China
Abstract:Objective To evaluated the performance of domestic LABSTAR automatic blood culture system comparison with BD BACTEC FX200 automatic blood culture system.Methods Two blood culture systems were used to simulate the specimens and clinical specimens in two parts,each of which was evaluated by a bottle and an anaerobic bottle.The standard strains and suspected or confirmed blood flow of clinical specimens were injected into their respective matching blood culture bottles were cultured,the comparative evaluation of domestic LABSTAR and the BD BACTEC FX200 results consistent rate,and positive results of the alarm time difference.Results Simulated specimens were evaluated:23 cases of simulated specimens wereevaluated by aerobic bottle group.All the two blood culture systems were reported positive and no significant difference was found between the two groups.Anaerobic bottle group to assess 14 cases of simulated samples,while the sun reported 11 cases were statistically significant difference.The results of clinical specimens:531 cases of oxygen bottles,245 cases of sun and the same results,the alarm time was statistically significant,the overall consistency rate of 96.4%.Anaerobic bottle group 361 cases,8 cases of sun and the results were the same,the alarm time was statistically different,and the overall consistency ratewas 95.8%.Conclusion The coincidence rate of DomesticLABSTAR automatic blood culture system and BD BACTEC FX200 automatic blood culture system was high,but the time of positive alarm time more long.
Keywords:
本文献已被 CNKI 等数据库收录!
点击此处可从《现代检验医学杂志》浏览原始摘要信息