Comparison of revision strategies for failed C2-posterior cervical pedicle screws: a biomechanical study
Authors:Michael Mayer  Juliane Zenner  Robert Bogner  Wolfgang Hitzl  Markus Figl  Arvind von Keudell  Daniel Stephan  Rainer Penzkofer  Peter Augat  Gundobert Korn  Herbert Resch  Heiko Koller
Affiliation:1. Department for Traumatology and Sports Injuries, Paracelsus Medical University, Muellner Hauptstrasse 48, 5020, Salzburg, Austria
2. German Scoliosis Center Bad Wildungen, Werner-Wicker-Klinik, Bad Wildungen, Germany
3. Research Office, Biostatistics, Paracelsus Medical University, Salzburg, Austria
4. Department for Traumatology, General Hospital, Tulln, Austria
5. Institute of Biomechanics, Trauma Center Murnau, Murnau, Germany
6. Institute of Biomechanics, Paracelsus Medical University, Salzburg, Austria
Abstract:

Study purpose

With increasing usage within challenging biomechanical constructs, failures of C2 posterior cervical pedicle screws (C2-pCPSs) will occur. The purpose of the study was therefore to investigate the biomechanical characteristics of two revision techniques after the failure of C2-pCPSs.

Materials and methods

Twelve human C2 vertebrae were tested in vitro in a biomechanical study to compare two strategies for revision screws after failure of C2-pCPSs. C2 pedicles were instrumented using unicortical 3.5-mm CPS bilaterally (Synapse/Synthes, Switzerland). Insertion accuracy was verified by fluoroscopy. C2 vertebrae were potted and fixed in an electromechanical testing machine with the screw axis coaxial to the pullout direction. Pullout testing was conducted with load and displacement data taken continuously. The peak load to failure was measured in newtons (N) and is reported as the pullout resistance (POR). After pullout, two revision strategies were tested in each vertebra. In Group-1, revision was performed with 4.0-mm C2-pCPSs. In Group-2, revision was performed with C2-pedicle bone-plastic combined with the use of a 4-mm C2-pCPSs. For the statistical analysis, the POR between screws was compared using absolute values (N) and the POR of the revision techniques normalized to that of the primary procedures (%).

Results

The POR of primary 3.5-mm CPSs was 1,140.5 ± 539.6 N for Group-1 and 1,007.7 ± 362.5 N for Group-2; the difference was not significant. In the revision setting, the POR in Group-1 was 705.8 ± 449.1 N, representing a reduction of 38.1 ± 32.9 % compared with that of primary screw fixation. For Group-2, the POR was 875.3 ± 367.9 N, representing a reduction of 13.1 ± 23.4 %. A statistical analysis showed a significantly higher POR for Group-2 compared with Group-1 (p = 0.02). Although the statistics showed a significantly reduced POR for both revision strategies compared with primary fixation (p < 0.001/p = 0.001), the loss of POR (in %) in Group-1 was significantly higher compared with the loss in Group-2 (p = 0.04).

Conclusions

Using a larger-diameter screw combined with the application of a pedicle bone-plastic, the POR can be significantly increased compared with the use of only an increased screw diameter.
Keywords:Cervical pedicle screw   Failure   Revision strategy   Biomechanics   Pedicle bone-plastic