Subannular perforation after long-lasting aortic valve replacement mimicking mitral insufficiency
Authors:Strauch Justus T  Wippermann Jens  Krep Henning  Wahlers Thorsten
Affiliation:Department of Cardiothoracic Surgery, University Hospital Cologne, Cologne, Germany. ju.strauch@gmx.de
Abstract:We report a rare case of a patient diagnosed with mitral insufficiency grade III 12 years after mechanical aortic valve replacement. Transesophageal echocardiography described an eccentric mitral regurgitation-type systolic jet with color flow evidence of communication between left ventricle and atrium. Surgical intervention showed a circular defect in the mitral-aortic intervalvular fibrosa area, after removal of the mechanical valve, located beneath the noncoronary sinus causing the echocardiography-detected mitral insufficiency. A pericardial patch was trimmed to the appropriate size, and the defect was closed. The aortic valve was replaced by a stented pericardial bioprosthesis.