Factors influencing radiation exposure during the extracorporeal shock wave lithotripsy |
| |
Authors: | W C Chen Y H Lee M T Chen J K Huang L S Chang |
| |
Affiliation: | Department of Surgery, National Yang-Ming Medical College, Taipei, Taiwan, Republic of China. |
| |
Abstract: | A prospective evaluation of 89 consecutive sessions of extracorporeal shock wave lithotripsy (ESWL) was undertaken to try and find the best way of minimising the amount of exposure to radiation. Forty-two patients were randomly allocated to undergo ESWL treatment by experienced surgeons (group A), and 47 to undergo the treatment by inexperienced surgeons (group B). The mean calculated entrance radiation exposure was 3.01 rads (group A: 2.64 (0.97) rads, range 1.00-4.48, group B: 3.38 (0.86) rads, range 1.11-5.75). Among factors that influenced radiation exposure, the tissue: air ratio should be borne in mind and the level of skill in controlling movement of gantry was the most important in reducing the exposure to radiation. |
| |
Keywords: | |
|
|