首页 | 本学科首页   官方微博 | 高级检索  
     


An evaluation of three commercial kits for use as screening methods for the detection of leptospiral antibodies in the UK
Authors:Zochowski W J  Palmer M F  Coleman T J
Affiliation:PHLS Leptospira Reference Unit/WHO/FAO Collaborating Centre for Research on Leptospirosis, County Hospital, Hereford HR1 2ER, UK. wendy.zochowski@talk21.com
Abstract:AIMS: To compare three commercial screening tests--the PanBio leptospiral IgM enzyme linked immunosorbent assay (ELISA), the Biolisa leptospiral IgM ELISA, and the indirect haemagglutination assay (IHA)--with the microscopic agglutination test (MAT) and two "in house" ELISAs--urease and horseradish peroxidase (HRP)--for the detection of leptospiral antibodies in a local UK and Eire population. METHOD: Two hundred sera submitted for a differential diagnosis of leptospirosis were tested by all methods. A further 142 sera from patients with antibodies to toxoplasma, Epstein-Barr virus (EBV), hepatitis A virus, rheumatoid factor, Borrelia burgdorferi, Mycoplasma pneumoniae, syphilis, cytomegalovirus, and Q fever were tested for crossreactivity. RESULTS: Compared with the MAT, sensitivity and specificity were found to be: PanBio, 90%/94%; Biolisa with sorbent, 100%/85%; and IHA, 54%/95%. Seven of 200 trial sera gave false negative results with PanBio; 14 of 200 trial sera gave false positive results with Biolisa with sorbent, as did a further 25 of the 142 sera tested for potential crossreactivity. Two of 142 sera gave crossreactions with PanBio and IHA (one each). CONCLUSIONS: The degree of false positivity seen with the Biolisa suggests that the recommended positive value of > or = 26 Eu/ml should be reassessed using pools of sera from local populations. When the cut off value was reassessed, using a value of > or = 40 Eu/ml, a sensitivity and specificity of 96% and 94%, respectively, was achieved. Even the modified Biolisa appears to be over sensitive and to show a high degree of non-specificity. The IHA, although specific (95%), lacked sensitivity in this study. The PanBio appeared to be the most suitable as a screening test for leptospiral IgM in the UK, although it would be advisable for all positive test results to be confirmed by a different enzyme immunoassay and the MAT.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号