首页 | 本学科首页   官方微博 | 高级检索  
     

高强度聚焦超声换能器声场的数值仿真与实验测量
引用本文:朱元光,张平,张晓静,菅喜岐.高强度聚焦超声换能器声场的数值仿真与实验测量[J].国际生物医学工程杂志,2011,34(2):78-81.
作者姓名:朱元光  张平  张晓静  菅喜岐
作者单位:天津医科大学生物医学工程学院,300070
基金项目:国家自然科学基金资助项目,天津市应用基础及前沿技术研究计划重点项目
摘    要:目的 高强度聚焦超声(HIFU)肿瘤治疗过程中,HIFU换能器形成的声场分布是决定其治疗效果的关键因素之一,为了提高HIFU肿瘤治疗的安全性和可靠性,需要对HIFU换能器形成的声场进行预测.方法 采用时域有限差分(FDTD)法数值仿真水体内形成的声压分布与实验测量结果对比的研究方法,分析讨论了HIFU换能器在不同激励功...

关 键 词:时域有限差分法  高强度聚焦超声换能器  声场

Simulation and measurement of acoustic field by HIFU transducer
ZHU Yuan-guang,ZHANG Ping,ZHANG Xiao-jing,JIAN Xi-qi.Simulation and measurement of acoustic field by HIFU transducer[J].International Journal of Biomedical Engineering,2011,34(2):78-81.
Authors:ZHU Yuan-guang  ZHANG Ping  ZHANG Xiao-jing  JIAN Xi-qi
Affiliation:. (School of Biomedical Engineering, Tianjin Medical University, Tianjin 300070, China)
Abstract:Objective In the treatment of tumor with high intensity focused ultrasound (HIFU),the acoustic pressure distribution determines the safety and reliability of HIFU.It is necessary to predict the acoustic field produced by HIFU transducer.Methods To analyze the acoustic pressure distribution of HIFU generated by different excitation frequency and variable incentive power,comparison was made between the acoustic pressure simulated by finite difference time domain (FDTD) method and the ones measured by experiments.Results The resultsshowed that the pressureat focalpoint increasedwiththe increasing of excitation power,while the change in the focal length was negligible.The focal point moved towards the transducer with the frequency increasing until the maximum sound pressure was gained at the resonant frequency (RF).Conclusion The simulation results are consistent with the experiment results,indicating that the acoustic fields can be predicted by simulation.
Keywords:Finite difference time domain method  High intensity focused ultrasound transducer  Acoustic field
本文献已被 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏