首页 | 本学科首页   官方微博 | 高级检索  
     


Influence of the remelting process on the fatigue behavior of electron beam irradiated UHMWPE
Authors:Puértolas J A  Medel F J  Cegoñino J  Gomez-Barrena E  Ríos R
Affiliation:Department of Material Science and Technology, Centro Politécnico Superior-I3A, Universidad de Zaragoza, E-50018, Zaragoza, Spain. japr@unizar.es
Abstract:Electron beam irradiation at doses below 150 kGy is a widely used technique to obtain highly crosslinked ultra-high-molecular-weight polyethylene (UHMWPE). Its current use in total joint replacement components may improve wear resistance and decrease UHMWPE particle debris. However, currently used post-irradiation thermal treatments, which aim to decrease the free radicals within the material, introduce microstructural changes that affect UHMWPE mechanical properties, particularly the fatigue strength. This influence may be crucial in total knee replacements, where fatigue-related damage limits the lifespan of the prosthesis. Therefore, more studies are required to understand UHMWPE fatigue after current crosslinking protocols. This study was planned to evaluate the influence of UHMWPE remelting after irradiation on the material fatigue resistance. The remelting was achieved at 150 degrees C for 2 h on UHMWPE previously irradiated at 50, 100, and 150 kGy. Fatigue evaluation included short-term tests under cyclic tensile stress with zero load ratio, R = 0, and 1 Hz. In addition, stress-life testing was performed using 12% yield as the criterion for failure. Near-threshold fatigue crack propagation experiments were also performed at a frequency of 5 Hz, and crack length was measured in nonthermally treated and remelted irradiated UHMWPE. Crystallinity percentage was calculated from DSC measurements. The results pointed out that irradiation positively contributed to total life analysis, but the further remelting process decreased the flaw initiation resistance. On the other hand, both processes negatively affected the fatigue resistance of notched components. From a clinical point of view, the results suggest that the material fatigue behavior should be carefully studied in new UHMWPE to avoid changes related to material processing.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号