MR imaging of the knee with a 0.064-T permanent magnet |
| |
Authors: | P A Rothschild J M Domesek L Kaufman D M Kramer S F Dye L J Anderson J L Lewis M K Gon |
| |
Affiliation: | Radiologic Imaging Laboratory, University of California, San Francisco 94080. |
| |
Abstract: | Magnetic resonance (MR) imaging has proved beneficial in the evaluation of internal derangements of the knee. A limitation to general acceptance of MR imaging of the knee has been availability and cost. The recent introduction of low-field-strength MR imaging has shown promise in decreasing the cost and increasing the availability of this modality. High-resolution (pixel size, 0.7 mm2), three-dimensional Fourier transform (3DFT), thin-section (3.5 mm) imaging performed on a 0.064-T permanent magnet was used to evaluate 117 knees in 114 consecutive patients. The appearance of normal anatomy and internal derangements of the knee at low-field-strength imaging is described. Arthroscopic correlation was available for 28 knees. Findings from low-field-strength MR imaging and arthroscopy agreed in 79% of cases in the determination of meniscal tears. Partial-flip-angle techniques with 3DFT produced thin-section images of the knee of diagnostic quality. The authors conclude that in patients with internal derangements of the knee, low-field-strength (0.064-T) MR imaging may provide useful information. |
| |