首页 | 本学科首页   官方微博 | 高级检索  
     


Efficacy and Safety of Permanent Cardiac DDD Pacing with Contemporaneous Double Spinal Cord Stimulation
Authors:MASSIMO ROMANO  STEFANIA BRUSA    ANTONIO GRIECO  FURIO ZUCCO    ACHILLE SPINELLI  BIAGIO ALLARIA
Affiliation:Divisione di Cardiologia, Servizio di Anestesia, Ospedale Santa Corona, Garbagnate Milanese, Italy;Rianimazione e Terapia del Dolore, Ospedale Santa Corona, Garbagnate Milanese, Italy
Abstract:Spinal cord stimulation (SCS) is currently used to treat peripheral vascular disease (PVD) and refractory angina pectoris not amenable to revascularization. In a case of contemporaneous SCS implant and permanent cardiac pacemaker (PPM), if multipolar electrodes are used it is possible to avoid any interference between the systems. We describe the case of a patient with a DDD pacemaker, in whom two bipolar SCSs were implanted at different times: one to control refractory angina pectoris and the other for PVD. No interference between the three systems has been observed.
Keywords:spinal cord stimulation    permanent cardiac pacemaker    electromagnetic interference