首页 | 本学科首页   官方微博 | 高级检索  
检索        


Noninvasive detection of left ventricular dysfunction with a portable electrocardiographic gated scintillation probe device
Authors:Arnold Strashun  Steven F Horowitz  Stanley J Goldsmith  Louis E Teichholz  Adam Dicker  Kenneth Miceli  Richard Gorlin
Institution:From the Andre Meyer Department of Physics-Nuclear Medicine and the Division of Cardiology and Nuclear Medicine, Department of Medicine, Mount Sinai Medical Center, New York, New York USA
Abstract:A comparison of left ventricular function data derived from a low cost, portable electrocardiographic gated scintillation probe (nuclear stethoscope) with conventional scintiangiographic data was performed in 68 patients. Ejection fraction correlation (r = 0.86, p < 0.005) was better in patients with uniform wall motion than in those with regional asynergy (r = 0.68 p < 0.01). Probe variables reflecting systolic emptying rates, diastolic filling rates and timing intervals, and relative volumes analyzed in combination provided 100 percent sensitivity, specificity, and predictive value in detecting abnormal left ventricular performance. The results suggest that radionuclide angiography with an electrocardiographic gated scintillation probe is a sensitive, rapid and relatively inexpensive portable method of screening for cardiac dysfunction with a yield similar to that from the more costly gamma camera derived scintiangiogram.
Keywords:Address for reprints: Arnold Strashun  MD  Department of Physics and Nuclear Medicine  Mount Sinai Medical Center  1 Gustave L  Levy Place  New York  New York 10029  
本文献已被 ScienceDirect 等数据库收录!