Detection of Doppler embolic signals: psychoacoustic considerations |
| |
Authors: | Chung Emma Fan Lingke Degg Chris Evans David H |
| |
Affiliation: | Department of Medical Physics, University Hospitals of Leicester, Leicester, UK. |
| |
Abstract: | The purpose of this study was to improve reliability in the identification of Doppler embolic signals by determining the decibel threshold for reproducible detection of simulated "emboli" as a function of signal duration, frequency and cardiac-cycle position. The auditory sensitivity of 16 participants to 574 simulated "emboli" was examined using psychoacoustic techniques to assess how the probability of detection varies with embolic signal parameters. Detailed measurements of the threshold for detection of simulated embolic signals are presented. These provide evidence that the measured embolus-to-blood threshold ranges between 2 dB and 14 dB as a continuous function of signal duration and frequency. The level of the threshold is closely linked to both embolic signal parameters and the properties of the blood flow signal. We conclude that the current fixed choice of threshold does not provide a good approximation to the true threshold of detection across the full range of embolic signal parameters. |
| |
Keywords: | |