首页 | 本学科首页   官方微博 | 高级检索  
     

耳石功能检查仪的设计与临床初步应用
摘    要:

关 键 词:耳石功能检查仪 设计 临床应用 前庭系统功能检查

Design and preliminary use of the device of subjective visual vertical]
Zi-ming Wu,Su-zhen Zhang,Cheng-jun Zhao,Wei-yan Yang,Dong-yi Han. Design and preliminary use of the device of subjective visual vertical][J]. Chinese Journal of Otorhinolaryngology, 2004, 39(9): 566-568
Authors:Zi-ming Wu  Su-zhen Zhang  Cheng-jun Zhao  Wei-yan Yang  Dong-yi Han
Affiliation:Department of Otorhinolaryngology Head & Neck Surgery, Otolaryngology Institute of General Hospital of Chinese PLA, Beijing 100853, China. zimingwu@163.com
Abstract:OBJECTIVE: The device of subjective visual vertical (SVV) was made to test the gravity orientation of subjects visual. METHODS: The normative value was established firstly in the right, left and both eyes respectively. Then the test was applied to evaluate the function of otolith function (utricular function) in normal adult subjects and examined the SVV in some patients suffered from acute unilateral vestibulopathy. RESULTS: The difference of the SVV among right, left and both eyes was not significant and the normative value was lower than +/- 2 degrees. No matter one eye or both eyes, SVV had the tendency to lateraliseze to the right. CONCLUSIONS: The SVV is a reliable choice to test the otolith vestibulopathy, which is at least suitable for acute phase. The bilateral eyes open mode is recommended to test the subjects suspected of utricuar dysfunction.