首页 | 本学科首页   官方微博 | 高级检索  
     


Quantitative and qualitative comparison of 3.0T and 1.5T MR imaging of the liver in patients with diffuse parenchymal liver disease
Authors:Tsurusaki Masakatsu  Semelka Richard C  Zapparoli Mauricio  Elias Jorge  Altun Ersan  Pamuklar Ertan  Sugimura Kazuro
Affiliation:a Department of Radiology, University of North Carolina at Chapel Hill, USA
b Department of Radiology, Graduate School of Medicine, Kobe University, Kobe, Japan
c Department of Radiology, School of Medicine of Ribeirao Preto, University of Sao Paulo, Ribeirao Preto, Brazil
Abstract:

Purpose

The purpose of our study was to compare signal characteristics and image qualities of MR imaging at 3.0 T and 1.5 T in patients with diffuse parenchymal liver disease.

Materials and methods

25 consecutive patients with diffuse parenchymal liver disease underwent abdominal MR imaging at both 3.0 T and 1.5 T within a 6-month interval. A retrospective study was conducted to obtain quantitative and qualitative data from both 3.0 T and 1.5 T MRI. Quantitative image analysis was performed by measuring the signal-to-noise ratios (SNRs) and the contrast-to-noise ratios (CNRs) by the Students t-test. Qualitative image analysis was assessed by grading each sequence on a 3- and 4-point scale, regarding the presence of artifacts and image quality, respectively. Statistical analysis consisted of the Wilcoxon signed-rank test.

Results

the mean SNRs and CNRs of the liver parenchyma and the portal vein were significantly higher at 3.0 T than at 1.5 T on portal and equilibrium phases of volumetric interpolated breath-hold examination (VIBE) images (P < 0.05). The mean SNRs were significantly higher at 3.0 T than at 1.5 T on T1-weighted spoiled gradient echo (SGE) images (P < 0.05). However, there were no significantly differences on T2-weighted short-inversion-time inversion recovery (STIR) images. Overall image qualities of the 1.5 T non-contrast T1- and T2-weighted sequences were significantly better than 3.0 T (P < 0.01). In contrast, overall image quality of the 3.0 T post-gadolinium VIBE sequence was significantly better than 1.5 T (P < 0.01).

Conclusions

MR imaging of post-gadolinium VIBE sequence at 3.0 T has quantitative and qualitative advantages of evaluating for diffuse parenchymal liver disease.
Keywords:Liver MRI   Chronic liver disease   MRI
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号