首页 | 本学科首页   官方微博 | 高级检索  
     


Correction of megavoltage cone-beam CT images for dose calculation in the head and neck region
Authors:Aubry Jean-François  Pouliot Jean  Beaulieu Luc
Affiliation:Department of Radiation Oncology, University of California San Francisco, 1600 Divisadero Street, Suite H1031, San Francisco, California 94143, USA. ambryjt@rodone.usef.edu
Abstract:Megavoltage cone-beam computed tomography (MVCBCT) imaging systems are now available for image-guided radiation therapy delivery and verification. In order to use the three-dimensional anatomical information for dose calculation, the MVCBCT image must provide accurate electron density. This work proposes a new method that has been developed to correct for the cupping and missing data artifacts seen on MVCBCT images of the head and neck region. It uses a conventional kilovoltage CT (kVCT) image as a reference for electron density and rigid registration with a MVCBCT image to obtain correction factors. Dose calculations performed on MVCBCT images corrected with the proposed method agree with calculations done on kVCT images within +/- 1% on phantoms. With patients images the agreement is within +/- 13% above the shoulders and +/- 5% below the shoulder line. This level of dose calculation accuracy allows the use of MVCBCT images for dose verification purposes.
Keywords:
本文献已被 PubMed 等数据库收录!