基于雷泰医疗多模态智能直线加速器VenusX的验收测试
引用本文:王海洋,解传滨,戴相昆,申红峰,陈高翔,俞伟,刘彦立,徐寿平,曲宝林. 基于雷泰医疗多模态智能直线加速器VenusX的验收测试[J]. 中国医疗设备, 2021, 0(4): 65-69
作者姓名:王海洋  解传滨  戴相昆  申红峰  陈高翔  俞伟  刘彦立  徐寿平  曲宝林
作者单位:中国人民解放军总医院第一医学中心放射治疗科
基金项目:国家重点研发计划(2017YFC0112100);国家自然科学基金项目(81801799);军队保健课题(20BJZ40)。
摘    要:目的 对雷泰医疗新型推出的医用直线加速器VenusX的各项技术指标进行临床验收测试,评估其临床应用的精确性、有效性及安全性.方法 参考国家相关标准规定与方法,对VenusX的辐射安全、机械精度、射束性能及影像系统进行质量检测与评估.结果 各项技术指标均符合相关标准要求:辐射安全项目正常有效;等中心大小0.5 mm、机架...

关 键 词:国产放疗设备  剂量特性  机械性能  验收测试

Acceptance Testing Based on LinaTech Multi-modal Intelligent Linear Accelerator VenusX
WANG Haiyang,XIE Chuanbin,DAI Xiangkun,SHEN Hongfeng,CHEN Gaoxiang,YU Wei,LIU Yanli,XU Shouping,QU Baolin. Acceptance Testing Based on LinaTech Multi-modal Intelligent Linear Accelerator VenusX[J]. Chinese medical equipment, 2021, 0(4): 65-69
Authors:WANG Haiyang  XIE Chuanbin  DAI Xiangkun  SHEN Hongfeng  CHEN Gaoxiang  YU Wei  LIU Yanli  XU Shouping  QU Baolin
Affiliation:(Department of Radiotherapy,The First Medical Center of Chinese PLA General Hospital,Beijing 100853,China)
Abstract:Objective To evaluate the accuracy,effectiveness and safety of Venusx,a new medical linear accelerator launched by LinaTech Medical,a clinical acceptance test was conducted on the technical indicators.Methods With reference to the relevant national standard regulations and methods,quality inspection and evaluation of radiation safety,mechanical accuracy,beam performance and imaging system of VenusX were conducted.Results All technical indicators meet the requirements of relevant standards:Radiation safety items were normal and effective.The size of the isocenter was 0.5 mm.The rotation angle of the gantry was 0.3°.The accuracy of the digital light distance scale was 1mm.The rotation angle of the treatment couch was 0.5°.The deviation of each direction of the couch was ≤1 mm.The accuracy and repeatability of MLC were better than the standard.The dose stability was better than the relevant standards.The center placement accuracy of MV-EPID was 0.4 mm,the placement repeatability accuracy was 0.2 mm,and the positioning accuracy was 0.5 mm.The center placement accuracy of kV-EPID was 0.3 mm,the placement repeatability accuracy was 0.2 mm,and the positioning accuracy was 0.1 mm.Conclusion As a domestic multimodal intelligent medical linear accelerator,the parameters and indicators of Venus X are better than the acceptance criteria,which proves that the safety,accuracy and effectiveness of the device are in line with relevant national standards and meet clinical requirements.
Keywords:domestic radiotherapy equipment  dose characteristics  mechanical properties  acceptance test